We have developed a photon detection system for muonic X-ray spectroscopy. The detector system consists of high-purity germanium detectors with BGO Compton suppressors. The signals from the detectors are readout with a digital acquisition system. The absolute energy accuracy, energy and timing resolutions, photo-peak efficiency, the performance of the Compton suppressor, and high count rate durability are studied with standard $\gamma$-ray sources and in-beam experiment using $^{ 27}\mathrm{ Al}(p, \gamma){ }^{ 28}\mathrm{ Si}$ resonance reaction. The detection system was demonstrated at Paul Scherrer Institute. A calibration method for a photon detector at a muon facility using muonic X-rays of $^{ 197}$Au and $^{ 209}$Bi is proposed.

Categories: physics.ins-det nucl-ex

Development of wide range photon detection system for muonic X-ray spectroscopy