Desorption electrospray ionization (DESI) is an ambient ionization technique that can be coupled to mass spectrometry (MS) for chemical analysis of samples at atmospheric conditions. Coupled ionization sources-MS systems are popular in chemical analysis because the individual capabilities of various sources combined with different MS systems allow for chemical determinations of samples. DESI employs a fast-moving charged solvent stream, at an angle relative to the sample surface, to extract analytes from the surfaces and propel the secondary ions toward the mass analyzer. This tandem technique can be used to analyze forensics analyses, pharmaceuticals, plant tissues, fruits, intact biological tissues, enzyme-substrate complexes, metabolites and polymers. Therefore, DESI-MS may be applied in a wide variety of sectors including food and drug administration, pharmaceuticals, environmental monitoring, and biotechnology.
History
DESI has been widely studied since its inception in 2004 by Zoltan Takáts, Justin Wiseman and Bogdan Gologan, in Graham Cooks' group from Purdue University with the goal of looking into methods that didn't require the sample to be inside of a vacuum. Both DESI and direct analysis in real time (DART) have been largely responsible for the rapid growth in ambient ionization techniques, with a proliferation of more than eighty new techniques being found today. These methods allow for complex systems to be analyzed without preparation and throughputs as high as 45 samples a minute. DESI is a combination of popular techniques, such as, electrospray ionization and surface desorption techniques. Electrospray ionization with mass spectrometry was reported by Malcolm Dole in 1968, but John Bennett Fenn was awarded a nobel prize in chemistry for the development of ESI-MS in the late 1980s. Then in 1999, desorption of open surface and free matrix experiments were reported in the literature utilizing an experiment that was called desorption/ionization on silicon. The combination of these two advancements led to the introduction of DESI and DART as the main ambient ionization techniques that would later become multiple different techniques. One in particular, due to increasing studies into optimization of DESI, is nanospray desorption electrospray ionization (nano-DESI). In this technique the analyte is desorbed into a liquid bridge formed between two capillaries and the analysis surface.

Principle of operation
DESI is a combination of electrospray (ESI) and desorption (DI) ionization methods. Ionization takes place by directing an electrically charged mist to the sample surface that is a few millimeters away. The electrospray mist is pneumatically directed at the sample where subsequent splashed droplets carry desorbed, ionized analytes. After ionization, the ions travel through air into the atmospheric pressure interface which is connected to the mass spectrometer. DESI is a technique that allows for ambient ionization of a trace sample at atmospheric pressure, with little sample preparation. DESI can be used to investigate in situ, secondary metabolites specifically looking at both spatial and temporal distributions.

Ionization mechanism
In DESI there are two kinds of ionization mechanism, one that applies to low molecular weight molecules and another to high molecular weight molecules. High molecular weight molecules, such as proteins and peptides show electrospray like spectra where multiply charged ions are observed. This suggests desorption of the analyte, where multiple charges in the droplet can easily be transferred to the analyte. The charged droplet hits the sample, spreads over a diameter greater than its original diameter, dissolves the protein and rebounces. The droplets travel to the mass spectrometer inlet and are further desolvated. The solvent typically used for the electrospray is a combination of methanol and water.

For the low molecular weight molecules, ionization occurs by charge transfer: an electron or a proton. There are three possibilities for the charge transfer. First, charge transfer between a solvent ion and an analyte on the surface. Second, charge transfer between a gas phase ion and analyte on the surface; in this case the solvent ion is evaporated before reaching the sample surface. This is achieved when the spray to surface distance is large. Third, charge transfer between a gas phase ion and a gas phase analyte molecule. This occurs when a sample has a high vapour pressure.
The ionization mechanism of low molecular weight molecules in DESI is similar to DART's ionization mechanism, in that there is a charge transfer that occurs in the gas phase.